advanced customized characterization technology

 
Product World Thin Film & Memory Products
TF 2000

  Specifications

Computer hard- and software:
  • Pentium III, IBM PC-compatible computer
  • VGA graphical interface
  • USB port
  • CD / DVD writer
  • 40 GB hard disk or larger
  • 128 MB RAM
  • Operating system Windows 2000
  • aixACCT's sophisticated and highly flexible ferroelectric test software
  • Remote and script control (optional)

FE module

Driving unit:
  • Voltage range ±25 V
    (optional external amplifier up to 10.000 V)
  • Output impedance 10 Ω
  • Maximum hysteresis excitation frequency (load dependent) < 1 MHz*
  • Min. pulse width 25 μs
  • Min. rise time 10 ns
  • Slew rate (typical) > 200 V/μs
  • Maximum capacitive load (freq. dependent) 1 μF
  • Maximum output current ±1 A

Current amplifier:
  • Voltage virtual ground input
  • Current range 1 pA - 1 A
  • High-voltage protection
  • Rise time ranges:
      Standard Enhanced*
    10 pA 10 ms 10 ms
    1 nA 900 μs 900 μs
    1 μA 50 μs 300 μs
    1 mA 5 μs 65 ns
    1 A 3 μs 25 ns
* min. values, with probe head adapted to sample

Fatigue parameter:
  • Maximum frequency (special amplifier) 20 MHz
    Test conditions: amplitude: 6 V peak to peak
      capacitive load: 10 pF

PS module

Driving unit:
  • Max. amplitude 10 V
  • Minimum rise time (at ohmic load) 1,2 ns
  • Repetition rate 100 kHz
  • Output resistance 70 Ω

Return:
Shunt resistance 100 Ω

Additional requirements:
A fast digital sampling oscilloscope with two active probe heads.

Oscilloscope data:
  • Analog bandwidth 1 GHz
  • Max. sampling rate 5 GS/s
  • IEEE interface

Sample specification:
  • Input capacitance 0,8 pF
  • Input resistance 1 MΩ

RX module

Voltage-step generator:
  • Max. amplitude ±10 V
  • Rise times 100 ns
  • Output resistance < 5Ω

Current amplifier:
  • Range 100 mA - 10 pA
  • Bandwidth (I > 50 μA) 300 kHz, (I > 1 mA) 500 kHz

DR module

Voltage-pulse generator:
  • Maximum amplitude ±10 V
  • Minimum pulse length 50 ns
  • Maximum pulse length 640 μs

Switching properties:
  • Turn-on delay time < 2 ns
  • Ron < 50 Ω
  • Roff > 10 13 Ω

High impedance electrometer:
  • Input resistance 1013 Ω
  • Input capacitance 10 pF
TF 2000
Main Page
Features
Specifications
Product Sheet/ Documentation (PDF)
Quick Links
Contact
Sales Partners